Radio Frequency Plasma Discharge Lamps for Use as Stable Calibration Light Sources
TBMG-15266
12/01/2012
- Content
Stable high radiance in visible and near-ultraviolet wavelengths is desirable for radiometric calibration sources. In this work, newly available electrodeless radio-frequency (RF) driven plasma light sources were combined with research-grade, low-noise power supplies and coupled to an integrating sphere to produce a uniform radiance source. The stock light sources consist of a 28 VDC power supply, RF driver, and a resonant RF cavity. The RF cavity includes a small bulb with a fill gas that is ionized by the electric field and emits light. This assembly is known as the emitter. The RF driver supplies a source of RF energy to the emitter.
- Citation
- "Radio Frequency Plasma Discharge Lamps for Use as Stable Calibration Light Sources," Mobility Engineering, December 1, 2012.